Chapter 5 Multi-Unit Commands
Calibrating the Libraries in a Multi-Unit Configuration
Quantum P4000 and P7000 Libraries User’s Guide 89
Calibrating the Libraries in a Multi-Unit Configuration 5
The
Calibrate All PTMs command enables you to calibrate all pass through
mechanisms in all libraries within the multi-unit configuration. Calibrate
the libraries during the initial installation and after any maintenance
procedure.
To calibrate all library PTM’s:
1 From the Master library in the multi-unit configuration, press the
Multi-Unit tab.
The GUI displays the
Multi-Unit screen (see figure 52).
2 Press the
Calibrate All PTMs button.
The GUI displays a
Command In Progress dialog box while the
calibration process is in progress.
Figure 52 Multi-Unit
Screen
Note: All libraries must be in standby mode.
Note: Multi-Unit calibration MUST be executed from the master
library designated during multi-unit configuration.
Multi-Unit
System
Off-line
Standby
Load Port
Operator Service
Multi
Unit
Back Forward
Home
O
Configure
Multi-Unit
Calibrate
All PTMs
Stop
About