Service Menu 4-139
September 2007
Fiducial Test Dialog
Path: Main Menu Service Diags Scanner Fiducial
Test
Use the Fiducial Test dialog to test the scanner for proper
operation. See Figure 4-209.
Figure 4-209 Fiducial Test Dialog
Parameters Description
Test 1 reads the Insert/Eject Station
1 fiducial and displays the
found edge coordinates.
2 reads the Insert/Eject
Station 2 and displays the
found edge coordinates.
Xcnt Displays the x edge position
count in tenths of a mm
Ycnt Displays the Y edge position
count in tenths of a mm
Continue Y to continue running the test.
N to cancel the test
Run this test several times to ensure the X and Y position
counts are within 10 tenths of a mm each time the test is
run.
Loops:00 Test: 1
Xcnt:000000
Ycnt:000000
Continue: Y<