Service Menu 4-155
September 2007
Corner Moves
Path: Main Menu Service Diags Library Corner
Moves
Use Corner Moves to move cartridges from the I/E station to
each corner of each storage section for all media types. There
must be three cartridges of the same type, in a row, present in
the I/E station for the test to work properly. Additionally, the
test will only exercise the corners in the library that are empty.
See Figure 4-241.
Figure 4-241 Corner Moves Dialog
Cell Beginning and ending storage
locations
Status Shows the status of the test
(Running, Canceled)
Cancel Y to cancel the test
N to continue running the test
CORNER MOVE TEST
Cell: 01 A 2 10
Status: Running
Cancel: N
Note
If the operation
stops, the Cancel
parameter no
longer displays.