Quantum Scalar 10K Provozní pokyny Strana 220

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 324
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 219
4-142 Menus and Commands
6-00058-05 Rev A
Step Dialog
Path: Main Menu Service Diags Tower Step
Use the Step Dialog to move the tower through each position.
See Figure 4-216.
Figure 4-216 Home Dialog
Parameters Description
Continue Y to continue the test
N to cancel the test
If you are in SERVICE mode and Y is selected for the
Continue, the following display appears and the tower is
ready to position. See Figure 4-217.
If you are not in SERVICE mode and Y is selected for
Continue, the WRONG TOWER MODE display
appears, select Accept to change to SERVICE mode.
See Figure 4-218.
Figure 4-217 Position Tower
Parameters Description
Tower Tower to be moved
(01–14, 8mm, DLT, LTO,
HALF)
Column Column number
01A–12L for DLT, LTO, HALF
01A–16P for 8mm
Amount The step width depending on
the media type
This test should be
used by trained
personnel ONLY!
Continue: N <
Position Tower
Tower: 01 LTO
Column:UNKNOWN <
Amount: 01
Zobrazit stránku 219
1 2 ... 215 216 217 218 219 220 221 222 223 224 225 ... 323 324

Komentáře k této Příručce

Žádné komentáře